LOINC display name : Troponin I.cardiac High sensitivity method [Mass/Vol];
LOINC long common name : Troponin I.cardiac [Mass/volume] in Serum or Plasma by High sensitivity method;
LOINC short name : Troponin I SerPl HS-mCnc;
LOINC description : The high sensitivity troponin assays manufactured by different vendors have different
analytical characteristics, including the limit of detection and 99% upper reference
limit (URL). The quantitative results from different assays should not be compared
to make clinical decisions, and reporting the quantitative result in isolation is
not recommended. The troponin high sensitivity quantitative results should be reported
with information about the assay and/or the interpretation, depending on the specific
reporting method used in a given laboratory or hospital, that are included in the
High sensitivity Troponin I [LOINC: 89577-1] and Troponin T [LOINC: 89576-3] panels.
1. The laboratory device 99th percentile upper reference limit [LOINC: 89581-3] 2.
The laboratory device coefficient of variation at 99th percentile URL [LOINC: 89580-5]
3. The laboratory device detection limit [LOINC: 87706-8] 4. Interpretation for troponin
T [LOINC: 89575-5] and/or troponin I [LOINC: 89578-9] [PMID: 25204966];